WHEN ACCURATE MEASUREMENT MATTERS.
Calibration – Each measuring instrument shows a certain measurement deviation. The degree of this deviation is influenced by various factors, such as the ambient temperature, for example. Therefore it’s necessary to periodically calibrate ultrasonic flowmeters, level meters, wall thickness meters, energy meters etc. and, in some cases, to adjust them.
When to calibrate?
Standards and legislation can stipulate that measuring instruments must demonstrably comply by means of calibration. In all other cases it is up to you to determine whether calibration is desirable.
Tip: When a measurement deviation endangers the quality of your flow processes, it is recommended to have your flowmeter calibrated periodically.
How often?
This question cannot be answered unambiguously. Decisive are the application of the flowmeter, the frequency with which the flowmeter is used and the specific properties of the instrument.
We recommend determining the calibration period per measuring instrument individually. Usually the calibration applies for the duration of one year. However, if highly accurate measurements are needed, the measurement uncertainty may be covered for a shorter term (e.g. 90 days).
After a number of calibrations, the course of the measurement deviation is usually easy to estimate. You can then decide to extend the term (for example 1 year, 2 years or 3 years). However, this method is not a guarantee.
Calibration Report
In order to demonstrate that measurements – performed with a specific flowmeter – are correct, it is good to record the calibration results in a report.
N.B. In the prospect of legal proceedings, it is increasingly the case that measurement results are presented as evidence (for example due to environmental legislation). If measurement results need to be above all suspicion, a calibration report can be very useful.